Gas Types: (C) Carrier; (D) Detector; (M) Make up; (N) Nebulizer; (P) Purge
Recommended Equipment: RL = REDLINE®, BL = BASELINE® regulators
Instrumentation | Impurity concern |
Types |
Recommended gas | Recommended equipment | |||
Flame Ionization Detector (FID) |
THC |
C/M: | HiQ Helium 5.0 |
BL C106 series |
|||
H2O | C/M: |
HiQ Nitrogen 5.0 |
BL C106 series | ||||
O2 | C/D: | HiQ Hydrogen 5.0 |
BL C106 series | ||||
D: | HiQ Synthetic Air 5.0 | BL C106 series | |||||
D: | 40% Hydrogen in Helium | BL C106 series | |||||
D: | 40% Hydrogen in Nitrogen | BL C106 series | |||||
Thermal Conductivity Detector (TCD) |
THC | C/M: |
HiQ Helium 5.0 | BL C106 series | |||
H2O | C/M: | HiQ Hydrogen 5.0 | BL C106 series | ||||
O2 | C/M: | HiQ Nitrogen 5.0 | BL C106 series | ||||
Electron Capture Detector (ECD) |
THC | C: | HiQ Helium 5.5 ECD | RL C200 series | |||
H2O | C: | HiQ Hydrogen 5.0 Zero | RL C200 series | ||||
O2 | C/M: | HiQ Nitrogen 5.0 ECD | RL C200 series | ||||
Halocarbons | C/M: | 5% Methane in Argon ECD | RL C200 series | ||||
C/M: | 10% Methane in Argon ECD | RL C200 series | |||||
Flame Photometric Detector (FPD) |
Sulfur or phosphorous components |
C: | HiQ Argon 5.0 Zero | BL C106 series | |||
(H2S, SO2, THT, sulfides) CO2 |
C: | HiQ Helium 5.0 Zero | BL C106 series | ||||
C/M: | HiQ Nitrogen 5.0 Zero | BL C106 series | |||||
C/D: | HiQ Hydrogen 5.0 Zero | BL C106 series | |||||
D: | HiQ Synthetic Air 5.0 | BL C106 series | |||||
Nitrogen Phosphorus Detector (NPD) |
NH3 | C/M: | HiQ Helium 5.0 | BL C106 series | |||
Amines | C/M: | HiQ Nitrogen 5.0 | BL C106 series | ||||
D: | HiQ Hydrogen 5.0 | BL C106 series | |||||
D: | HiQ Synthetic Air 5.0 | BL C106 series | |||||
Photo Ionization Detector (PID) |
THC | C: | HiQ Helium 5.0 Zero | BL C106 series | |||
CO2 | C: | HiQ Nitrogen 5.0 Zero | BL C106 series | ||||
Pulse Discharge Photo Ionization Detector (PDPID) |
C/P: | HiQ Helium 6.0 | RL C200 series | ||||
Supercritical Fluid Chromatography (SFC) |
H2O non-volatile; liquid phase FID- & ECD-responsive; gas phase; O2 & THC |
C: | HiQ Carbon dioxide 4.5 SFC | BL C106 series | |||
High-Performance Liquid Chromatography (HPLC) |
P: | HiQ Helium 5.0 | BL C106 series | ||||
|
|||||||
Evaporative Light Scattering Detector (ELSD) | N: | HiQ Nitrogen 4.6 | BL C106 series |
* The gas recommendations include minimum purity requirement.